Model Number 376297
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/18/2019
Type of Reportable Event Malfunction
Event or Problem Description
Passeo-18 balloon catheters were chosen for treatment.During the same treatment it was not possible to cross the lesion with three passeo-18 balloons with different sizes.Finally a rotablator was used.
 
Additional Manufacturer Narrative
Neither the complaint instrument nor the angiographic material was provided for analysis.Therefore no technical investigation on the subject could be performed.The production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause was determined.The final root cause is most likely related to the patients anatomy (severe calcification).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PASSEO-18 2.5/200/130
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key8896338
Report Number1028232-2019-03466
Device Sequence Number1561262
Product Code LIT
UDI-Device Identifier07640130430044
UDI-Public07640130430044
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K151744
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2021
Device Model Number376297
Device Catalogue NumberSEE MODEL NO.
Device Lot Number01190142
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/25/2019
Supplement Date Received by Manufacturer10/10/2019
Initial Report FDA Received Date08/15/2019
Supplement Report FDA Received Date10/14/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
-
-