• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AGA MEDICAL CORPORATION AMPLATZER TORQVUE DELIVERY SYSTEM; CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AGA MEDICAL CORPORATION AMPLATZER TORQVUE DELIVERY SYSTEM; CATHETER, PERCUTANEOUS Back to Search Results
Model Number 9-ITV06F180/80
Medical Device Problem Codes Retraction Problem (1536); Material Twisted/Bent (2981)
Health Effect - Clinical Codes Low Blood Pressure/ Hypotension (1914); Complete Heart Block (2627)
Date of Event 07/18/2019
Type of Reportable Event Serious Injury
Event or Problem Description
Related manufacturer report number : 2135147-2019-00255.On (b)(6) 2019, an amplatzer muscular vsd occluder was selected for implant.The device was deployed but didn't fit the ventricular septal defect.The device was attempted to be retracted, but the occluder didn't collapse and retract into the delivery sheath.The tip of the delivery sheath kept folding in on itself.Meanwhile, the patient went into complete heart block and became hypotensive.The occluder and delivery system were withdrawn from the patient without collapsing the occluder.The procedure was cancelled and the patient was sent to the pediatric intensive care unit for continued monitoring.The heart block resolved without treatment.
 
Additional Manufacturer Narrative
The reported event of an inverted sheath tip was confirmed.Gross morphological examination revealed the tip of the sheath was bent and deformed, consistent with the soft tip inverting.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications at the time of commercialization.The cause of the reported event could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER TORQVUE DELIVERY SYSTEM
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
MDR Report Key8897452
Report Number2135147-2019-00256
Device Sequence Number259631
Product Code DQY
UDI-Device Identifier00811806010779
UDI-Public00811806010779
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K072313
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2022
Device Model Number9-ITV06F180/80
Device Catalogue Number9-ITV06F180/80
Device Lot Number7003532
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/28/2019
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 07/19/2019
Supplement Date Received by Manufacturer09/13/2019
Initial Report FDA Received Date08/15/2019
Supplement Report FDA Received Date10/08/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age5 YR
-
-