The reason for this revision surgery was due to a planned staged reverse.The previous surgery and the surgery detailed in this investigation occurred 5.9 months apart.The healthcare professional indicated there was a significant adverse event to the patient.There was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr), shows that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There were no nonconforming material reports associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to planned staged reverse.There were no findings during this investigation that indicate that the reported device was defective.The surgeon performed this procedure to remedy the patient's condition.No further no information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
|