| Catalog Number |
PAPS2 |
| Medical Device Problem Codes |
Suction Problem (2170); Noise, Audible (3273)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/21/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has been returned and the investigation results are pending.A follow up mdr will be submitted upon completion of the device investigation.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a thrombectomy procedure of the acute cerebral using a pump max canister (canister).During the procedure, the physician noticed the aspiration pressure would only reach -10 inhg and the canister made sound of leaking air.Therefore, the physician removed the canister and the procedure was completed using a new canister.There was no report of an adverse effect to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Results: the returned canister was undamaged.Conclusions: evaluation of the returned pump max canister (canister) revealed a functional device.During the functional testing, the canister was tested with a demonstration pump max and able to produce a vacuum pressure within specification.The reported air leaking sound from the canister could not be confirmed.Penumbra canisters are 100% inspected and functionally tested during incoming quality inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
| |
|