| Model Number |
G18344 |
| Medical Device Problem Codes |
Premature Activation (1484); Failure to Advance (2524)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); Foreign Body In Patient (2687)
|
| Date of Event |
07/23/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation evaluation: the user returned the product pouch.The lot number on the pouch matches the lot number in the report.A product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.The instructions for use also states: "with clip closed and without holding handle spool, advance device in small increments into accessory channel of gastroscope.Caution: holding handle spool during advancement may prematurely deploy clip." prior to distribution, all instinct endoscopic hemoclips are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
| |
|
Event or Problem Description
|
|
During a hemostasis procedure, the physician used an instinct endoscopic hemoclip.The clip exited the endoscope two seconds before the sheath [clip prematurely deploys in endoscope channel].The sheath was not saved and the clip is floating inside the patient.The following additional information was provided on 02-aug-2019: the closed clip deployed while being advanced through the endoscope channel.The prematurely deployed clip remained in the patient¿s body to pass naturally.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
| |
|
Search Alerts/Recalls
|