| Model Number |
CRPLUS |
| Medical Device Problem Codes |
Failure to Power Up (1476); Power Problem (3010)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
07/23/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Physio-control advised the customer that their device is not field-serviceable and is no longer under warranty.Physio recommended that the customer remove the device from service and a replacement device be obtained.The device was not returned to physio- control for evaluation.The cause of the reported issue could not be determined.Device not evaluated by manufacturer.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device was showing all three icons (attention, charge-pak and service wrench) and the device will not power on.With all three icons illuminated, the device may not have sufficient power available to deliver effective defibrillation therapy to a patient, if it was needed.There was no patient use associated with the reported event.
|
| |
|
Search Alerts/Recalls
|