Model Number 50-501XL
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Uterine Perforation (2121)
Date of Event 07/29/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device has not yet been returned therefore, a failure analysis of the complaint device cannot be completed.If the device is returned and evaluation completed, a supplemental medwatch will be filed device history record (dhr) review was conducted for the reported identification number.The lot was released meeting all qa specifications.(b)(4).
 
Event or Problem Description
It was reported that during the procedure, a scope was used to visualize the cavity.It appeared there were two holes at the fundus of the uterus but they were believed to be scar tissue.The deficit was maintaining and not rising.The tissue removal device was inserted into the cavity, visualization was lost and the fluid deficit spiked.The device was removed.Abdominal and vaginal ultrasounds were performed and there was no fluid found in patient abdomen.The device was discarded and the procedure was aborted.Patient was being watched in observation room.No additional details available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MYOSURE XL HYSTEROSCOPIC TISSUE REMOVAL SYSTEM
Common Device Name
UTERINE TISSUE REMOVAL SYSTEM
Manufacturer (Section D)
HOLOGIC, INC
250 campus drive
marlborough MA 01752
MDR Report Key8899477
Report Number1222780-2019-00188
Device Sequence Number240865
Product Code HIH
UDI-Device Identifier35420045505098
UDI-Public(01)35420045505098(10)
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K122498
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/06/2021
Device Model Number50-501XL
Device Catalogue Number50-501XL
Device Lot Number18J06RA
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/29/2019
Initial Report FDA Received Date08/15/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-