CODMAN AND SHURTLEFF, INC OG TDL CMPLX FILL COIL 4X12; NEUROVASCULAR EMBOLIZATION DEVICE Back to Search Results
Catalog Number 640CF0412
Medical Device Problem Code Separation Failure (2547)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/31/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturer¿s ref.No: (b)(4).Information regarding patient identifier, date of birth, age, gender, weight, race, ethnicity, and medical history were not provided.Procode is: krd/hcg.(b)(6).The initial reporter email address is not available.Conclusion: the healthcare professional reported that during the coil embolization procedure targeting an unruptured aneurysm, the 4mm x 12cm orbit galaxy complex fill coil (640cf0412 / 30029281) was used but the coil could not be detached even though several attempts were made to detach the coil.The coil was successfully removed from the patient.It was reported that the coil delivery system was prepped without any difficulty.The coil was replaced with a competitive product and the procedure was completed.The reported event did not result in any interruption of blood flow.There was no report of patient injury / complication or procedural delay as a result of the reported event.Based on complaint information, the device was not available to be returned for analysis.A review of manufacturing documentation associated with this lot (30029281) presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no non-conformances related to device manufacture or inspection.All product rejected during manufacturing was identified as scrap and properly accounted for.Product analysis cannot be conducted as the product was not returned for analysis.No determination of causes and possible contributing factors could be made.As such, the investigation will be closed.Failure to detach is a known potential issue associated with the use of this device.The instructions for use (ifu) contains several precautions related to this issue and includes instructions for troubleshooting the situation should it be encountered during use.With the limited information available and without the product available for analysis, the reported customer complaint could not be confirmed.Based on the device history record review, there is no indication that the event is related to the device manufacturing process.The exact cause of the event could not be conclusively determined; however, it is possible that circumstances of the procedure and/or device manipulation/interaction may have contributed to the reported failure.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective/preventive action may be triggered at a later time.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.This is one of 2 products involved with the reported complaint.The associated manufacturer report number is 3008114965-2019-01145.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
 
Event or Problem Description
The healthcare professional reported that during the coil embolization procedure targeting an unruptured aneurysm, the 4mm x 12cm orbit galaxy complex fill coil (640cf0412 / 30029281) was used but the coil could not be detached even though several attempts were made to detach the coil.The coil was successfully removed from the patient.It was reported that the coil delivery system was prepped without any difficulty.The coil was replaced with a competitive product and the procedure was completed.The reported event did not result in any interruption of blood flow.There was no report of patient injury / complication or procedural delay as a result of the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OG TDL CMPLX FILL COIL 4X12
Common Device Name
NEUROVASCULAR EMBOLIZATION DEVICE
Manufacturer (Section D)
CODMAN AND SHURTLEFF, INC
325 paramount dr
raynham MA 02767
MDR Report Key8899535
Report Number1226348-2019-00976
Device Sequence Number19236748
Product Code KRD
UDI-Device Identifier10886704030270
UDI-Public10886704030270
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K093973
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2020
Device Catalogue Number640CF0412
Device Lot Number30029281
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/31/2019
Initial Report FDA Received Date08/15/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-