Manufacturer¿s ref.No: (b)(4).Information regarding patient identifier, date of birth, age, gender, weight, race, ethnicity, and medical history were not provided.Procode is: krd/hcg.(b)(6).The initial reporter email address is not available.Conclusion: the healthcare professional reported that during the coil embolization procedure targeting an unruptured aneurysm, the 4mm x 12cm orbit galaxy complex fill coil (640cf0412 / 30029281) was used but the coil could not be detached even though several attempts were made to detach the coil.The coil was successfully removed from the patient.It was reported that the coil delivery system was prepped without any difficulty.The coil was replaced with a competitive product and the procedure was completed.The reported event did not result in any interruption of blood flow.There was no report of patient injury / complication or procedural delay as a result of the reported event.Based on complaint information, the device was not available to be returned for analysis.A review of manufacturing documentation associated with this lot (30029281) presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no non-conformances related to device manufacture or inspection.All product rejected during manufacturing was identified as scrap and properly accounted for.Product analysis cannot be conducted as the product was not returned for analysis.No determination of causes and possible contributing factors could be made.As such, the investigation will be closed.Failure to detach is a known potential issue associated with the use of this device.The instructions for use (ifu) contains several precautions related to this issue and includes instructions for troubleshooting the situation should it be encountered during use.With the limited information available and without the product available for analysis, the reported customer complaint could not be confirmed.Based on the device history record review, there is no indication that the event is related to the device manufacturing process.The exact cause of the event could not be conclusively determined; however, it is possible that circumstances of the procedure and/or device manipulation/interaction may have contributed to the reported failure.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective/preventive action may be triggered at a later time.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.This is one of 2 products involved with the reported complaint.The associated manufacturer report number is 3008114965-2019-01145.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
|