Model Number TRX-310
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 07/17/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.The reason for reoperation: capsular contracture baker grade iv.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.This is a known potential adverse event addressed in the product labeling.
 
Event or Problem Description
Healthcare professional reported left side capsular contracture, baker grade iv.Device remains implanted.
 
Additional Manufacturer Narrative
Analysis of the returned device identified the weight of the device was within specification, deformation and creases fold.Cloudy and voids in the gel observed after autoclave cycle.Based on the device analysis, the final assessment is no issues found related with the manufacturing process.
 
Event or Problem Description
Patient reported left side "hard, calcification and pain".Healthcare professional additionally reported patient's concern with the product.Device has been explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
CS 
MDR Report Key8900546
Report Number9617229-2019-10435
Device Sequence Number253978
Product Code FTR
UDI-Device Identifier10888628008472
UDI-Public10888628008472
Combination Product (Y/N)N
Initial Reporter StateNV
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2020
Reporter Type Manufacturer
Report Source consumer,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/16/2021
Device Model NumberTRX-310
Device Catalogue NumberTRX-310
Device Lot Number2885029
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/11/2020
Initial Date Received by Manufacturer 07/17/2019
Supplement Date Received by Manufacturer03/11/2020
Initial Report FDA Received Date08/16/2019
Supplement Report FDA Received Date04/10/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age59 YR
-
-