MAUDE Adverse Event Report: MEDIVANCE, INC. ¿ 1725056 ARCTIC SUN TEMPERATURE MANAGEMENT SYSTEM, ARCTICGEL PADS; ARCTIC GEL PAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDIVANCE, INC. ¿ 1725056 ARCTIC SUN TEMPERATURE MANAGEMENT SYSTEM, ARCTICGEL PADS; ARCTIC GEL PAD Back to Search Results
Catalog Number 318-02
Device Problems Break (1069); Inaccurate Flow Rate (1249); Temperature Problem (3022)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 07/25/2019
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event Description
It was reported that the flow rate was 0.5 l / min on the arctic sun device and the patient's temperature could not be controlled.Also, the tube inside the pad was found to be broken.
 
Manufacturer Narrative
The reported event was unconfirmed.Visual evaluation of the returned sample noted one opened (no original packaging present), arctic gel neonatal pad present.Visual inspection of the pad surface noted no obvious visible defects, such as a cut or tear in the foam.Visual inspection of the clear connectors noted no visible chips and deformities at the ends of all four connectors.A kink was observed in the tubing.The tubing was not broken.The flow rate was found to be acceptable for the returned pad.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: ¿directions 1.Place the patient (1.8 - 4.5 kg; 4.0 - 9.9 lb) on the pad.Avoid placing the patients over the manifolds or other high pressure locations.The rate of temperature change and potentially the final achievable temperature is affected by pad surface area coverage, placement, patient size, and water temperature range.2.The pad surface must be contacting the skin for optimal energy transfer efficiency.A) if desired, the center section of the pad can be wrapped around the patient¿s torso and secured in place using the velcro tabs provided.¿ if this option is in use, ensure that the edges of the pad are away from articulating areas of the body to avoid irritation.¿ place pads to allow for full respiratory excursion.(e.G.Ensure free movement of the chest and abdomen are guaranteed).¿ the pads may be removed and reapplied if necessary.¿ pads should be placed on healthy, clean skin only.3.Due to the small patient size (1.8 - 4.5 kg; 4.0 - 9.9 lb) and the potential for rapid patient temperature change, it is recommended to use the following settings to the arctic sun® temperature management system: ¿ water temperature high limit: =40°c (104°f) ¿ water temperature low limit: =10°c (50°f) ¿ control strategy: 2 4.Due to the small patient size (1.8 - 4.5 kg; 4.0 - 9.9 lb) it is recommended to use the patient temperature high and patient temperature low alert settings.5.Place a core patient temperature probe and connect to the arctic sun® temperature management system patient temperature 1 connector for continuous patient temperature feedback.A rectal or esophageal temperature probe is recommended.6.Verify patient core temperature with an independent temperature probe before and at regular intervals during use.7.Attach the pad¿s line connectors to the fluid delivery line manifolds.8.See arctic sun® temperature management system operators manual and help screens for detailed instructions on system use.9.Begin treating the patient.10.If the pad fails to prime or a significant continuous air leak is observed in the pad return line, check the connections, then if needed, replace the leaking pad.Once the pad is primed, assure the steady state flow rate displayed on the control panel is appropriate.The minimum flow rate should be 1.1 l/m.11.When finished, purge water from pad.".
 
Event Description
It was reported that the flow rate was 0.5 l / min on the arctic sun device and the patient's temperature could not be controlled.Also, the tube inside the pad was found to be broken.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCTIC SUN TEMPERATURE MANAGEMENT SYSTEM, ARCTICGEL PADS
Type of Device
ARCTIC GEL PAD
Manufacturer (Section D)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
louisville CO 80027
MDR Report Key8901058
MDR Text Key154869713
Report Number1018233-2019-04793
Device Sequence Number1
Product Code DWJ
Combination Product (y/n)N
PMA/PMN Number
K142702
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,other
Type of Report Initial,Followup
Report Date 09/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/16/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number318-02
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/28/2019
Date Manufacturer Received09/04/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-