• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TULSA DENTAL PRODUCTS LLC ASEPTICO 30K ELECTRIC MOTOR HANDPIECE, DIRECT DRIVE, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TULSA DENTAL PRODUCTS LLC ASEPTICO 30K ELECTRIC MOTOR HANDPIECE, DIRECT DRIVE, AC-POWERED Back to Search Results
Catalog Number AM4B
Device Problems Mechanical Problem (1384); Energy Output Problem (1431)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/18/2019
Event Type  malfunction  
Manufacturer Narrative
There has been a previous report received where this malfunction has caused file separation. Since separation of a file could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur. As such, this event meets the criteria for reportability per 21 cfr part 803. The device is available for evaluation, though has not been returned as of this report. Evaluation results will be submitted as they become available.
 
Event Description
It was reported that an aseptico m4b motor was running slower than usual; no injury resulted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameASEPTICO 30K ELECTRIC MOTOR
Type of DeviceHANDPIECE, DIRECT DRIVE, AC-POWERED
Manufacturer (Section D)
TULSA DENTAL PRODUCTS LLC
608 rolling hills drive
johnson city TN 37604
Manufacturer (Section G)
ASEPTICO, INC
8333 216th st se
woodinville WA 98072
Manufacturer Contact
karl nittinger
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494424
MDR Report Key8901196
MDR Text Key191030968
Report Number2320721-2019-00162
Device Sequence Number1
Product Code EKX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K030163
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 10/03/2019
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received08/16/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue NumberAM4B
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received09/30/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-