| Model Number |
1884004 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Radiation Exposure, Unintended (3164)
|
| Date of Event |
07/22/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Analysis found that the distal tip of the inner cutter broke off which would have resulted in the reported event.The breakage occurred at the first proximal tooth valley.The portion that broke off measured approximately 0.24¿ in length.The outer assemblies were deformed in such a way that indicates the inner blade teeth impacted the outer teeth at the 1st proximal tooth while moving in a ccw direction which resulted in the observed break.There were no signs of misuse or mishandling however there was some gouging on the inner shaft outside diameter 0.56¿ from the distal face of the inner hub proceeding to 1.05¿ from the hub.Deformation of the outer tube distal tip has been proven to cause the inner blade tip to fracture at the first proximal tooth valley.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
A health care provider (hcp) reported via manufacturer representative that during a bilateral reduction of gynaecomastia using liposuction and handpiece.The handpiece was used to remove the glandular breast tissue.During the procedure the blade broke off and had to be retrieved with x-ray and extra incision was made to retrieve the broken part.The procedure was delayed for forty five minutes.The procedure was completed with backup product.
|
| |
|