| Model Number |
29462 |
| Medical Device Problem Codes |
Material Integrity Problem (2978); Separation Problem (4043)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
07/13/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by the manufacturer: the device was returned for analysis.The tip and inner/outer shaft were microscopically and visually inspected.Inspection revealed the outer shaft was separated at the strain relief, and there was a kink in the shaft located 4mm from the strain relief.
|
| |
|
Event or Problem Description
|
|
It was reported that the tip of the sheath was disconnected with the tube.A renegade hi-flo fathom system was selected for used.During preparation, when unpacking the plastic package, it was noted that a disconnection from the tube of the tip of the sheath and could not be used.The procedure was completed with another of the same device and the patient is stable.No complications reported.
|
| |
|