| Model Number |
MMT-751RNAL |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/11/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Customer reported via phone call that the insulin pump had unexpected restart alarm.The customer¿s blood glucose level was 356 mg/dl at the time of the incident.Customer reported they were unable to clear the alarm.Customer able to complete insulin pump rewind.Customer stated unexpected restart alarm was reoccurred.Troubleshooting was performed.Customer was advised that the insulin pump needed to be replaced.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Insulin pump received with blank display due to corroded battery tube spring contact and corroded battery tube.Unable to perform all functional testing including the operating currents, a21 error test, rewind, basic occlusion, occlusion, prime and excessive no delivery test or verify a21 alarm due to blank display.
|
| |
|