| Model Number |
MMT-7008B |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/10/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call the sensor had inaccurate readings that triggered threshold suspend alarm.Insulin delivery was suspended due to sensor values.Sensor glucose value that triggered suspend on low or suspend before low event was 60 mg/dl.Low limit in sensor settings was 70.The blood glucose value associated with the triggered suspend event 169 mg/dl.The customer was informed that their blood glucose and sensor glucose levels were not in acceptable range.The sensor will be returned for analysis.
|
| |
|