MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA COIL 400; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA COIL 400; HCG, KRD Back to Search Results
Catalog Number 4002C0720
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Ischemia (1942)
Date of Event 12/19/2013
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.Potential adverse events in the labeling with the penumbra system include, but are not limited to, hematoma or hemorrhage at the site, inability to completely remove thrombus, intracranial hemorrhage, ischemia, including death.Therefore, it was determined that the reported adverse events were anticipated complications.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2015-00855.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the internal carotid artery using penumbra coil 400 (pc400) coils.The procedure was successful with no complications.However, the next day, magnetic resonance imaging revealed small foci of silent ischemia in the left cerebral hemisphere of the patient.The relationship of the pc400 coils to the ischemia is uncertain; however the relationship of the procedure to the ischemia is definite.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA COIL 400
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key8905884
Report Number3005168196-2019-01583
Device Sequence Number19515747
Product Code HCG
UDI-Device Identifier00814548010540
UDI-Public00814548010540
Combination Product (Y/N)Y
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K120330
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/30/2015
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2018
Device Catalogue Number4002C0720
Device Lot NumberF38122
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/30/2015
Initial Report FDA Received Date08/19/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age47 YR
Date Report Sent to FDA01/01/2005
-
-