Catalog Number 305601
Medical Device Problem Code Component Missing (2306)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/26/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Oem manufacture: the manufacturing location for this product is flextronics.This site is an oem manufacturing site.Therefore, bd corporate headquarters in (b)(4) has been listed and the (b)(4) fda registration number has been used for the manufacture report number.
 
Event or Problem Description
It was reported that the sharps coll 9gal gasket red experienced a missing lid or slide clamp which was noted prior to use.The following information was provided by the initial reporter: material no.305601, batch no.9033926.It was reported the lid was missing.
 
Additional Manufacturer Narrative
H.6.Investigation summary: it was reported that lids were missing.A sample and photo representation was provided for evaluation.A review of the device history record (dhr) for the reported lot was performed and there were no issues found for missing lids.A review of non-conforming material report (ncmr) was performed for the last twelve months and did not identify any nonconformance reported for missing lids for the same part number.This lot was confirmed produced after corrective actions.A weight system that provides a weighted label for each final box before shipment has already been implemented for this product to ensure the correct quantity of pieces per box before shipping.It was not possible to confirm that any failure mode was related to the manufacturing process with the information provided since the weight system is validated and did not detect any missing components.There may have potentially been a repackaging issue which occurred during distribution.The actual root cause is unknown.Based on these findings, our investigation is inconclusive.Bd will continue to monitor for any trends.
 
Event or Problem Description
It was reported that the sharps coll 9gal gasket red experienced a missing lid or slide clamp which was noted prior to use.The following information was provided by the initial reporter: material no.305601; batch no.9033926.It was reported the lid was missing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHARPS COLL 9GAL GASKET RED
Common Device Name
HYPODERMIC SINGLE LUMEN NEEDLE
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
MDR Report Key8906265
Report Number2243072-2019-01752
Device Sequence Number19317238
Product Code FMI
UDI-Device Identifier50382903056014
UDI-Public50382903056014
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K943575
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number305601
Device Lot Number9033926
Initial Date Received by Manufacturer 07/26/2019
Supplement Date Received by Manufacturer07/26/2019
Initial Report FDA Received Date08/19/2019
Supplement Report FDA Received Date09/12/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-