A steris service technician arrived onsite to inspect the table and found the retaining tab of the fuse holder within the ac plate assembly to be broken off.The damaged ac plate caused an inadequate electrical connection resulting in the reported event.The table was installed in 1998 making it approximately 21 years old and is not under steris service agreement.The user facility is responsible for all maintenance activities.The 3085sp surgical table operator manual states (1-2), "warning - personal injury and/or equipment damage hazard: safe and reliable operation of this equipment requires regularly scheduled preventive maintenance, in addition to the regular performance of routine maintenance.Contact steris to schedule preventive maintenance." the reported event is attributed to improper maintenance activities.The user facility did not perform proper preventive maintenance activities, specifically the biannual inspection of the ac plate, as outlined in the 3085sp maintenance manual.The 3085sp surgical table preventive maintenance check list states, "4.3 check power cord and ac plate for damage.Replace as needed.2 x per year." the technician replaced the ac plate, tested the table, confirmed it to be operating according to specification and returned it to service.The technician counseled facility personnel on the proper use and operation of the 3085sp surgical table, specifically performing proper preventive maintenance.No additional issues have been reported.
|