• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER SUSTAINABILITY SOLUTIONS; ARTHROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SUSTAINABILITY SOLUTIONS; ARTHROSCOPE Back to Search Results
Model Number 375-544-000
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/29/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to stryker sustainability solutions for evaluation.Visual inspection revealed no evidence of clinical use.The inner shaft tip was completely detached from the shaft and returned with the device.A review of the dhr for the reported lot number supports that the device met all inspection and test criteria prior to release from stryker.Therefore, the most likely root causes are the user applied too much force to the device or the arthroscopic shaver may have come into contact with staples, clips, or another metal object, resulting in damage to the blade.The instructions for use (ifu) state: before beginning the procedure, verify compatibility of all instruments and accessories.Plug in and set up the generator according to the instructions in the manufacturers¿ manual.Select an arthroscopic shaver with size, blade, and function most appropriate for the procedure.Inspect the instrument for overall condition and physical integrity.Do not use the instrument if any damage is noted.Return the instrument and packaging to stryker sustainability solutions if it is not in acceptable condition for surgery.Follow a suitable surgery protocol.Careful handling of the instrument is necessary to avoid damage or breakage as a result of excessive force.Do not apply excessive pressure or ¿side-load¿ the blade during use.Side-loading does not improve the performance of the instrument, can dull the blade, and/or produce metal particulates.Do not allow the arthroscopic shaver to come into contact with staples, clips or any metal object to avoid damage to the blade and possible patient injury.The tip of the bur or cutter must be irrigated periodically (general recommendation: once a minute) to cool the blade and prevent excised tissues from accumulating.Do not run the instrument without appropriate suction for the duration of the process.The reported event will continue to be monitored through post-market surveillance.
 
Event or Problem Description
It was reported the tip broke off the shaver during the knee scope.The complainant was not aware of any patient injury or medical intervention.The extended procedure time was just the time it took to retrieve the piece.The complainant was not aware of the length of extended procedure time.The tip fell into the patient and was retrieved with an arthroscopic grasper.These are commonly used devices that are readily available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
STRYKER SUSTAINABILITY SOLUTIONS
1810 w. drake drive
tempe AZ 85283
MDR Report Key8907973
Report Number0001056128-2019-00042
Device Sequence Number19444949
Product Code HRX
UDI-Device Identifier00885825014247
UDI-Public00885825014247
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K012635
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/16/2022
Device Model Number375-544-000
Device Catalogue Number375-544-000RR
Device Lot Number10059969
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/06/2019
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/29/2019
Initial Report FDA Received Date08/19/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
-
-