LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
|
Back to Search Results |
|
Model Number 16-02-80 |
Device Problem
Microbial Contamination of Device (2303)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 05/08/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no known patient involvement.(b)(6).The heater-cooler 16-02-80 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k052601).Livanova (b)(4) implemented a field safety notice for disinfection and cleaning of heater-cooler devices.The z number is z-2076/2081-2015.Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).A review of the dhr could not identify any deviations or nonconformities relevant to the issue.Corrective actions are in progress for this issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Device not returned.
|
|
Event Description
|
Livanova (b)(4) received a report, that a heater-cooler system 3t was found to be contaminated with mycobacterium chimaera.Hospital performed lab test, which confirms the contamination with mycobacterium chimaera.The report has been provided to livanova (b)(4).There is no known patient involvement.
|
|
Manufacturer Narrative
|
H.10: through follow-up communication livanova learned that the device was cleaned according to the instruction for use and that it was placed inside the operating theatre during use at an estimated distance from the surgery field of 2 meters, with the fan directed opposite to the patient.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
See initial report.
|
|