• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ATRIUM MEDICAL CORPORATION ADVANTA VXT W/GDS; PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ATRIUM MEDICAL CORPORATION ADVANTA VXT W/GDS; PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER Back to Search Results
Model Number 22114
Health Effect - Clinical Code Inflammation (1932)
Date of Event 08/08/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
On completion of the investigation a follow up report will be submitted.
 
Event or Problem Description
After graft transplantation, there was inflammation.The user assumed that the problem occurred elsewhere rather than the graft problem.
 
Additional Manufacturer Narrative
Analysis: the details indicate that swelling occurred after the procedure and if other devices from the same lot had been reported for swelling issues.A review of our complaint log indicates that there have been no other complaints related to swelling from the same production lot of graft going back historically 3 years.A thorough review of the device history records for this graft lot was conducted.The device lot in question met all quality and performance testing requirements and no non-conformances were noted during the build of the product related to the complaint.The instructions for use in the adverse reaction section mentions the following in regards to complication with the use of any vascular graft include the following: ¿complications that may occur in connection with the use of any vascular graft include, but are not limited to; thrombosis, stenosis, formation of pseudoaneurysm due to excessive needle punctures, peri-graft hematoma formation, peri-graft seroma formation, excessive needle hole bleeding or weeping, infection, swelling of tissue, suture hole elongation, mechanical disruption, material separation, delamination or tearing of the graft material, suture line or host vessel which may result in extreme blood loss, loss of limb function, steal syndrome, loss of limb or possible death.The patient should be advised to contact the physician should an adverse reaction occur.¿ conclusion: based on the review of the device history records and product complaint details atrium medical corporation cannot conclude that the swelling was directly related to the advanta graft.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANTA VXT W/GDS
Common Device Name
PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER
Manufacturer (Section D)
ATRIUM MEDICAL CORPORATION
40 continental blvd
merrimack NH 03054
MDR Report Key8910258
Report Number3011175548-2019-00884
Device Sequence Number525607
Product Code DSY
UDI-Device Identifier00650862221145
UDI-Public00650862221145
Combination Product (Y/N)N
Initial Reporter CountryKR
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/10/2022
Device Model Number22114
Device Catalogue Number22114
Device Lot Number443557
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Initial Date Received by Manufacturer 08/13/2019
Supplement Date Received by Manufacturer09/10/2019
02/01/2021
Initial Report FDA Received Date08/20/2019
Supplement Report FDA Received Date09/17/2019
02/05/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age84 YR
Patient Weight61
-
-