| Model Number |
N/A |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
No Information (3190); No Code Available (3191)
|
| Date of Event |
08/04/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Device name: spectrum turbo-ject double lumen minocycline / rifampin bedside power-injectable picc.A follow-up report will be submitted should additional relevant information become available.
|
| |
|
Event or Problem Description
|
|
It was reported that the line of a spectrum turbo-ject double lumen minocycline/rifampin bedside power-injectable picc was "broken on the lumen of the catheter." additional information has been requested with regards to the device, event and patient details but have not been provided at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation - evaluation: a review of documentation including the complaint history, device history record, instructions for use (ifu) and quality control of the device was conducted during the investigation.The complaint device was not returned; therefore, a physical examination could not be completed.However, a complaint device was returned for a previous complaint (mfg.Report reference #: 1820334-2019-01317).A separation was found below the white hub and extension tubing.The extension tubing lengths, outer diameters and the picc tube proximal outer diameter were measured and all found to be within manufacturing specifications.A definitive cause of the investigation was unable to be established.Additionally, a document-based investigation evaluation was performed.It was concluded that sufficient inspection activities are in place to identify this failure mode prior to device distribution.A review of the device history record for lot 9593532 found no nonconformances that could have contributed to the failure mode.It should be noted that there were no other complaints reported for this lot number.There is no evidence to suggest that nonconforming product exists in house or in the field.The instructions for use (ifu), provides the following information to the user related to the reported failure mode: intended use: "the maximum pressure limit setting for power injectors used with spectrum turbo-ject picc may not exceed 325 psi and the flow rate may not exceed the maximum flow rate indicated, as shown on the following table.4fr double lumen maximum flow rate for the orange hub is 3 ml/sec.Do not power inject the white hub lumen.Injection pressure limit setting is 325 psi." warnings: "do not power inject if maximum injection rate cannot be verified to meet limit printed on catheter hub or extension tube." precautions: "if lumen flow is impeded, do not force injection or withdrawal of fluids." catheter maintenance: "if microccave or other needleless adapters approved for saline only lock are used, saline only catheter lock may be used.Catheter heparinization should be determined by institutional protocol and clinical judgement.Heparin concentrations for 10 units/ml to 100 units/ml have been reported adequate to maintain lumen patency.Catheter lock should be reestablished after every use or at least every 24 hours if unused.¿ instructions for use: "use only lumens marked "ct" for power injection of contrast media.Warning: use of lumens not marked "ct" for power injection may result in catheter failure." "attach a 10 ml syringe filled with sterile normal saline to the hub of the extension tube to be used for power injection." "ensure adequate blood return and flush catheter vigorously with the entire 10 ml of sterile normal saline to ensure lumen patency.Warning: failure to ensure patency of the catheter lumen prior to injection may result in catheter failure." how supplied: "upon removal from package, inspect the product to ensure no damage has occurred." based on the information provided, no returned product and the results of our investigation, a definitive root cause could not be established.Appropriate measures have been taken to address this failure mode.The appropriate personnel have been notified and we will continue to monitor for similar complaints.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.Blank fields on this form indicate the information is unknown, unchanged or unavailable.
|
| |
|
Event or Problem Description
|
|
No new event description information to report at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: d10, h1, h3, h6: d10 - device available for eval: yes.The device was received on 29oct2019.H1 - reportable event type: serious injury h3 - device evaluated by mfg: yes, evaluation summary attached.H6 - method code: device was returned and tested.A review of documentation including the complaint history, device history record, instructions for use (ifu), quality control, and specifications of the device, as well as a dimensional verification and visual inspection were conducted during the investigation.The complaint device was received for evaluation on 29oct2019, along with a sticky note with additional procedural information.Upon inspection of the device, there was a gap found between the orange hub and extension tubing.A leak test was performed on the orange hub to find a leak between the hub and extension tubing.The extension tubing outer and inner diameter were measured to be within specification.The reported failure of the device being broken and leaking below the hub could be recreated.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Event or Problem Description
|
|
Additional information was received on 01oct2019 indicating that the device will be returned for investigation.Upon receipt of the device additional information was received stating that the complaint device was replaced with a competitor's catheter on (b)(6) 2019.
|
| |
|
Event or Problem Description
|
|
Additional information regarding patient and event information was received on 28aug2019.It was reported that during a picc line insertion in a vein in the arm on (b)(6) 2019, the picc line cracked and broke below the hub which resulted in oozing.The device had to be removed immediately on (b)(6) 2019 and replaced with a new one.No section of the device was left remaining inside the patient's body.Additional procedures were required due to this occurrence and the device.No adverse effects were reported due to this occurrence.
|
| |
|
Additional Manufacturer Narrative
|
|
H6 - patient code: no code available - additional surgical procedure to remove and replace the device.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Search Alerts/Recalls
|