| Catalog Number |
7D2649 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Information (3190)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Testing was performed at abbott diagnostics (b)(4).On retained kit lot 104656 with the following internal whole blood control samples: (b)(6).All test results were valid and performed as expected.Additionally, the manufacturing batch records for lot 104656 were reviewed.This lot met the required release specifications.A review of the complaints reported as false positive or unconfirmed (b)(6) related to lot number 104656 showed that the complaint rate is (b)(4).The evidence available does not indicate that the product is performing outside label claims.Abbott diagnostics (b)(4).Was unable to determine the exact root cause of the reported issue.The results obtained may possibly be related to the patient sample.The sample may have contained specific substances which may have affected the results.The available evidence suggests that this device lot is performing within labeled claims.
|
| |
|
Event or Problem Description
|
|
A customer reported (b)(6) results with the alere determine hiv-1/2 ag/ab combo test between (b)(6) 2019 and (b)(6) 2019.Confirmation testing with a 4th generation hiv test (not otherwise specified) was nonreactive for all six (6) patients.The sample type and specific result (antibody/antigen) was not reported.Patient gender, pregnancy status, treatment and patient outcome were not reported.There is insufficient information to determine if a malfunction occurred.Attempts to gain additional information were not successful.
|
| |
|
Search Alerts/Recalls
|