| Catalog Number |
201203040 |
| Medical Device Problem Codes |
Device-Device Incompatibility (2919); Scratched Material (3020)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Investigation summary: the device was reviewed and confirmed the connection part of the device is worn.Root cause attributed to the device being worn from normal use and servicing.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the tri-lock broach size 4 (p/n: unknown) and the broach handle were not able to connect during the tha surgery on (b)(6) 2019.As a result of confirming, the broach handle was able to connect with other all broaches, so there was a possibility of the broach in question had been distorted or had been scratched.The surgery was completed, and it was unknown whether there was a surgical delay or not.There was no adverse consequence to the patient.No further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|