The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.Based on the information provided, the reported difficulties appear to be due to operational circumstances of the procedure.The stent migration was the result of inadvertent interaction with the newly deployed stent and the tip of the delivery system.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.G9: exemption number e2019001-permits numbering sequence to begin with 10000, to avoid duplication of report numbers due to process transition.There may be gaps in numbering for reports submitted during the transition period.
|