ABBOTT VASCULAR ARMADA 35 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Catalog Number B2060-200
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/27/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
G9: exemption number e2019001-permits numbering sequence to begin with 10000, to avoid duplication of report numbers due to process transition.There may be gaps in numbering for reports submitted during the transition period.Visual analysis was performed on the returned device.The reported balloon rupture was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents.The investigation determined that the reported difficulties were likely due to case related circumstances.It is likely that the balloon interacted with the heavy lesion calcification causing damage to the outer surface of the balloon material which subsequently ruptured during inflation.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was to treat a superficial femoral artery.A 6.0 x 200 mm armada 35 balloon dilatation catheter ruptured during the first inflation at 14 atmospheres.The device was removed and replaced.There were no adverse patient effects and no clinically significant delay during the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key8913456
Report Number2024168-2019-11009
Device Sequence Number19269114
Product Code LIT
UDI-Device Identifier08717648155321
UDI-Public08717648155321
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K111899
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2022
Device Catalogue NumberB2060-200
Device Lot Number90327G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/07/2019
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/31/2019
Initial Report FDA Received Date08/20/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-