| Model Number |
ZXR00 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Blurred Vision (2137); No Code Available (3191)
|
| Date of Event |
07/31/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown, not provided, but the best estimate date is between (b)(6) 2019 and (b)(6) 2019.(b)(4).Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a zxr00 intraocular lens (iol) was explanted.There was no intervention performed.Additional information obtained revealed the lens was explanted from the patients right eye due to blurry vision.No additional information was received.
|
| |
|
Additional Manufacturer Narrative
|
|
Section d10: device available for evaluation: yes.Section d10: returned to manufacturer on: 8/19/2019.Section h3: device returned to manufacturer: yes.Device evaluation: the intraocular lens (iol) was returned to the manufacturing site for evaluation.Visual inspection shows the product is cut in pieces, most probably to make explant possible.Considering the condition of the lens additional analysis is not possible.The complaint cannot be confirmed.Manufacturing records review: the manufacturing records for the product was reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no similar complaints have been received for this production order number.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
|
| |
|
Search Alerts/Recalls
|