| Model Number |
MMT-723RNAS |
| Medical Device Problem Code |
Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/11/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump had a partial display.The customer¿s blood glucose level was unknown at the time of the incident.Customer stated that the blank screen on the insulin pump.Customer stated the black screen did not disappear.The device will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed self test and displacement test.Device display properly.No missing segment/partial display anomaly or full back screen noted during testing.
|
| |
|