| Catalog Number |
705004 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/26/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device will not be returned.If additional information becomes available, it will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
The customer reported that the drill cold welded with the locking drill guide during the procedure.There was a 5 minute delay to surgery.The customer further reported that the drill bit doesn't actually fit the sleeve/tower on the axsos 3 plating system.As a result of this mismatch, the surgeon believes that it explains why the bit becoming stuck in the sleeve if the direction is off by a couple of degrees.
|
| |
|
Search Alerts/Recalls
|