| Model Number |
MT22719 |
| Medical Device Problem Code |
Defective Alarm (1014)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/24/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that intermittent audio output occurred.Product has been received but is pending evaluation.A follow up report will be submitted upon completion.No injury or medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
The product was evaluated.An exterior visual inspection was performed and passed.Charge and boot testing were performed and passed.Receiver functional testing was performed and passed.G5 global communication tool tone at medium testing was performed and passed.Manual functional tests "try it" was performed and passed.Open receiver case for internal visual inspection was performed and passed.Measure the speaker resistance was performed and passed.The receiver log was downloaded and reviewed finding no errors related to the complaint.The allegation was not confirmed.The probable cause could not be determined.No injury or medical intervention was reported.
|
| |
|