• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number MT22719
Medical Device Problem Code Defective Alarm (1014)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/24/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that intermittent audio output occurred.Product has been received but is pending evaluation.A follow up report will be submitted upon completion.No injury or medical intervention was reported.
 
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
The product was evaluated.An exterior visual inspection was performed and passed.Charge and boot testing were performed and passed.Receiver functional testing was performed and passed.G5 global communication tool tone at medium testing was performed and passed.Manual functional tests "try it" was performed and passed.Open receiver case for internal visual inspection was performed and passed.Measure the speaker resistance was performed and passed.The receiver log was downloaded and reviewed finding no errors related to the complaint.The allegation was not confirmed.The probable cause could not be determined.No injury or medical intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM
Common Device Name
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
MDR Report Key8915938
Report Number3004753838-2019-061258
Device Sequence Number16597393
Product Code PQF
UDI-Device Identifier20386270000235
UDI-Public20386270000235
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P120005
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMT22719
Device Catalogue NumberSTK-SD-001
Device Lot Number5250439
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/20/2019
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 07/24/2019
Supplement Date Received by Manufacturer09/17/2019
Initial Report FDA Received Date08/21/2019
Supplement Report FDA Received Date10/01/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age62 YR
Patient Weight59
-
-