| Model Number |
FGS-0450 |
| Medical Device Problem Code |
Unintended Application Program Shut Down (4032)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/19/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is based on information provided by medtronic investigation personnel.The product sample was not returned to the medtronic laboratory; however, a picture was provided by the customer for analysis.The returned sample did not meet specification as received by medtronic.The customer reported they had a 96-hour study that ended early and only 47 minutes were recorded.The reported condition was confirmed.The investigation found that the reported condition was due to bravo system communication failure.The investigation isolated the failure to the bravo communication system, but a cause was not identified.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, they had a 96-hour study that ended early and only 47 minutes were recorded.A copy of the study was sent in for review.The recorder worked correctly during the previous procedure.There was no patient and user harm and a repeat procedure was necessary.
|
| |
|