| Model Number |
MS1-4570S |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
05/22/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
No device has been returned for evaluation nor have radiographs or photographic images been provided.Insufficient information has been provided to determine root cause.No additional investigation can be completed at this time.Labeling review ".Warning: metallic implants can loosen, fracture, corrode, migrate, or cause pain.".
|
| |
|
Event or Problem Description
|
|
As per reporter it is unknown if product may or may not have reached terminal distraction length.Patient underwent a revision procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
The rods were received for functional and visual testing and the reported event was confirmed.The root cause of the reported event may be related to bending force applied during the distraction sessions and/or patient's daily activities.
|
| |
|