CAREFUSION ALARIS PUMP MODULE ADMINISTRATION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 2420-0007
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/31/2019
Event Type  malfunction  
Manufacturer Narrative
The customer¿s report of being able to disconnect the closed system transfer device from the tubing was confirmed.Visual inspection of the set noted no damage or any anomalies.Functional and pressure testing resulted in the fluid flowing freely and showed no signs of leaking or occluding.The set was loaded into a lab pump module for an infusion test.The primary infusion was programmed at a rate of 125ml/h and 125 ml vtbi.The infusions completed successfully with no alarms and no anomalies throughout the set.The root cause of the customer¿s report was not identified.
 
Event Description
It was reported that the nurse was able to disconnect the closed system transfer device from the tubing despite being "clicked on" to the tubing, however the "c-clamped was not able to be placed on".The customer states there was no injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARIS PUMP MODULE ADMINISTRATION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer (Section G)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer Contact
ade ajibade
10020 pacific mesa blvd
san diego, CA 92121-4386
8586172000
MDR Report Key8919251
MDR Text Key155186901
Report Number9616066-2019-02395
Device Sequence Number1
Product Code FPA
UDI-Device Identifier07613203021012
UDI-Public7613203021012
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K944320
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Nurse
Type of Report Initial
Report Date 08/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/21/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/10/2022
Device Model Number2420-0007
Device Catalogue Number2420-0007
Device Lot Number19065648
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/07/2019
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/10/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
NON-BD SPIROS ADAPTER, THERAPY DATE (B)(6) 2019
-
-