| Model Number |
MMT-523RNAS |
| Medical Device Problem Code |
Computer Software Problem (1112)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/13/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump alarmed failure of flash memory error.The customer's blood glucose value was 150 mg/dl.Customer reported they were able to clear the alarm.Customer stated they were not able to rewind the pump.The customer was advised that the pump will need to be replaced and was advised to discontinue use of the pump and revert to a back-up plan.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received alarming a32 followed by an e22 alarm, problem isolated to mother board.Unable to perform self test and displacement test due to a32 and e22 alarm.
|
| |
|