| Model Number |
M0062201070 |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/05/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).(b)(4).Investigation results: visual examination of the returned complaint device found the balloon burst.No more damage found on the device.Functional analysis could not be performed due to the returned condition of the balloon.This failure is likely due to factors or conditions related to the procedure that could have affected its performance and its intended purpose such as; the manner as the device was handled, and/or the interaction with the scope during the procedure.Therefore, the most probable root cause is adverse event related to procedure.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.
|
| |
|
Event or Problem Description
|
|
Boston scientific corporation received an unauthorized return of an occluder balloon catheter that was used in the kidney during a percutaneous nephrolithotomy (pnl) procedure performed on (b)(6) 2019.It was found that the balloon ruptured during the procedure.Reportedly, the procedure was completed with another occluder balloon catheter.There were no patient complications reported as a result of the event.The patient condition at the conclusion of the procedure was reported to be good.
|
| |
|