| Catalog Number |
2100-12 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
07/22/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
We have received the complaint device for evaluation and we have confirmed the reported incident.When we attempted to inflate the balloon with saline, we observed a leakage at the joint of the inflation arm and the irrigation lumen.The root cause of the issue was determined to be a manufacturing operator error- not enough glue was applied on the joint during assembly process.Our review of the lot history records for this lot did not find any discrepancies either in the manufacturing or packaging process that could be related to this incident.Further, we have not received any other complaints of a similar nature for devices from this lot.Therefore, we believe that it was an isolated incident.The malfunction was detected by the user during pre-use check.Device was not used in the patient.Procedure was completed using another pruitt aortic catheter that they had in stock.
|
| |
|
Event or Problem Description
|
|
During pre-use check, surgeon detected a leakage at the joint where the inflation arm meets the irrigation lumen.Device was not used in the patient.Surgeon completed the procedure using another pruitt aortic occlusion catheter that they had in stock.
|
| |
|