• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SPINE INC XPDM QUICK-CON DI POLY SCWDRVR; SCREWDRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SPINE INC XPDM QUICK-CON DI POLY SCWDRVR; SCREWDRIVER Back to Search Results
Model Number 279722400
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/14/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device is available for evaluation.Investigation will be conducted.Follow up will be filed with the investigation results.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
The original plif procedure was performed to stabilize l4-5 approximately three years ago.It was reported that this was a revision procedure (plif) treating spinal stenosis to stabilize l3-s on (b)(6) 2019.After inserting a screw on the right l5, the surgeon moved on to screw insertion on the left l5 with the screwdriver (279722400).The screw gradually did not go deeper.Replacing a ratchet handle did not correct the issue.When the surgeon detached a qc driver from the unit, it was found that the screwdriver¿s tip had broken off at the screw¿s core section.The screw, stuck with the fragment, was removed from the patient¿s body.The surgeon completed the procedure with a replacing screw less than a 30-minute surgical delay.Not much load was put during the screw¿s insertion.No further information is available.Concomitant devices reported: unknown screw, (part# unknown, lot# unknown, quantity# unknown) this complaint involves one (1) device.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Udi: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Visual examination at the microscopic level revealed that the fracture was located at the driver¿s distal tip.The second half of the driver¿s tip was not returned.The fracture analysis report reveals plastic deformation at the hex lobes and torsional shear markings following a circular pattern.This suggests the fracture tip underwent a quasi-static overload torsional shear failure.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.A definitive root cause for the driver¿s distal tip becoming fractured cannot be positively determined.However, the fracture analysis report suggests that the fractured tip underwent a quasi-static overload torsional shear failure.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer, this complaint file will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XPDM QUICK-CON DI POLY SCWDRVR
Common Device Name
SCREWDRIVER
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
MDR Report Key8920654
Report Number1526439-2019-52019
Device Sequence Number19469995
Product Code HXX
UDI-Device Identifier10705034199344
UDI-Public(01)10705034199344
Combination Product (Y/N)N
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number279722400
Device Catalogue Number279722400
Device Lot NumberMI60388
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/20/2019
Initial Date Received by Manufacturer 08/14/2019
Supplement Date Received by Manufacturer09/09/2019
11/01/2019
Initial Report FDA Received Date08/22/2019
Supplement Report FDA Received Date10/01/2019
11/04/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-