| Catalog Number |
5531G111 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/26/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A review of the device history records indicate that devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Device not available.
|
| |
|
Event or Problem Description
|
|
It was reported by the sale rep that small gap was found in between tibia tray and tibia insert after the insert was log in to the tibia tray new tibia insert was used to completed the case.
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding size/fit issue involving a triathlon insert was reported.The event was confirmed based on the images provided.Method & results -product evaluation and results: visual inspection of the returned device noted the following: damage consistent with the implantation/explantation process was observed on the anterior region and articulating surface.Nothing noteworthy was observed on the surface of the locking wire.No materials or manufacturing discrepancies were observed on the surfaces examined.-medical records received and evaluation: no medical records were received for review with a clinical consultant.-product history review: a review of the device manufacturing records indicate that all devices accepted into final stock were free from discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: an intraoperative photograph was provided which confirms the gap between the insert and baseplate components.Visual inspection of the returned device noted the following: damage consistent with the implantation/explantation process was observed on the anterior region and articulating surface.Nothing noteworthy was observed on the surface of the locking wire.No materials or manufacturing discrepancies were observed on the surfaces examined.While the event was confirmed based on the provided images, the exact cause of the event could not be determined, it is not known if any intraoperative event took place which may have resulted in the seating/locking issue.No further investigation for this event is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Event or Problem Description
|
|
It was reported by the sale rep that small gap was found in between tibia tray and tibia insert after the insert was log in to the tibia tray new tibia insert was used to completed the case.
|
| |
|