| Model Number |
ASPC24-3E. |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
Blood Loss (2597)
|
| Date of Event |
08/09/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device involved was discarded by the facility and not returned for evaluation.Photographs were provided.The photograph revealed a slice in the lumen.The slice was reported to have been under the gauze dressing.The ballooning occurred after the lumen was clamped to prevent blood loss through the slice.As the device was implanted for approximately one year, the root cause of ballooning is most likely not due to a manufacture issue.Without an evaluation of the actual device, a definitive root cause cannot be determined.
|
| |
|
Event or Problem Description
|
|
After being implanted for approximately 12 months, the lumen of the catheter was noted to balloon.A slice/cut appears on a portion of the lumen which was under a gauze dressing.The ballooning was reported to occur after the catheter was clamped to block the blood from leaking out from the slice.The catheter was used only for dialysis and was not used for power injections.Catheter was removed.
|
| |
|