Model Number G34279
Medical Device Problem Code Difficult to Advance (2920)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/03/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Pma/510(k) # k160229.One unit of echo-hd-22-ebus-p-c of lot c1587797 was returned opened in its original packaging.A proximal kink was observed on the needle.Needle was able to advance and retract with slight difficulty.Stylet able to be removed and inserted without issue.Prior to distribution, all echo-hd-22-ebus-p-c devices are subjected to functional checks and visual inspection to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cook (b)(4), a review of the manufacturing records for echo-hd-22-ebus-p-c of lot number c1587797 did not reveal any discrepancies that could have contributed to this complaint issue.The review of relevant manufacturing records, confirms the failure mode has not previously occurred with the current lot number.Based on the information available to date, there is no evidence to suggest that there are any manufacturing issues associated with lot number c1587797.The notes section of the instructions for use, ifu0109-5 which accompanies this device instructs the user to; "visually inspect with particular attention to kinks, bends and breaks.If an abnormality is detected that would prohibit proper working condition, do not use" and ¿this device is intended for use with an olympus ebus scope" and "ensure the stylet is fully inserted when advancing the needle into the biopsy site".There is evidence to suggest that the customer did not follow the instructions for use in relation to type of scope used and the use of the stylet (ifu0109-5).A definitive root cause for the customer complaint could not be determined as circumstances of use cannot be replicated in the laboratory.A possible root cause could be attributed to user error as the stylet should not be partially removed prior to advancement of the needle into intended targeted site.Complaint is confirmed as the failure was verified in the laboratory.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Complaints of this nature will continue to be monitored for potential emerging trends.
 
Event or Problem Description
Needle is not smooth during procedure(difficult to push needle out from shealth), but tested outside scope was fine (when it is in straight position).Very hard for doctor to use the needle to do biopsy process while dr (b)(6) try another set of needle is fine.Additional information provided on 25-jul-19 confirmed that the stylet was removed about 2-3cm when advancing to target site.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHOTIP PROCORE ENDOBRONCHIAL HD BIOPSY NEEDLE
Common Device Name
FCG KIT, NEEDLE, BIOPSY
Manufacturer (Section D)
COOK IRELAND LTD
o halloran road
limerick
MDR Report Key8921159
Report Number3001845648-2019-00430
Device Sequence Number16743484
Product Code FCG
UDI-Device Identifier00827002342798
UDI-Public(01)00827002342798(17)220221(10)C1587797
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/21/2022
Device Model NumberG34279
Device Catalogue NumberECHO-HD-22-EBUS-P-C
Device Lot NumberC1587797
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/09/2019
Was the Report Sent to FDA? No
Distributor Facility Aware Date07/03/2019
Event Location Hospital
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/04/2019
Initial Report FDA Received Date08/22/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age62 YR
-
-