Catalog Number 71674068
Device Problem Break (1069)
Patient Problem No Information (3190)
Event Date 07/24/2019
Event Type  malfunction  
Event Description
It was reported that threads on drivers sheared when inserting.Backup was available and no delay was recorded.
 
Manufacturer Narrative
The associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The tip of the threads on the retaining rod fractured off, rendering the device inoperable.The fractured threads were not returned.Only one device was returned for evaluation.The device was manufactured in 2017 and exhibits signs of significant wear/usage.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUBTROC LAG SCREW DRIVER
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8921833
MDR Text Key155244403
Report Number1020279-2019-03131
Device Sequence Number1
Product Code HSB
UDI-Device Identifier03596010560612
UDI-Public03596010560612
Combination Product (y/n)N
PMA/PMN Number
K122170
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 10/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71674068
Device Lot Number12ASP0009
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/22/2019
Initial Date Manufacturer Received 07/24/2019
Initial Date FDA Received08/22/2019
Supplement Dates Manufacturer Received07/24/2019
Supplement Dates FDA Received10/20/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-