Additional Manufacturer Narrative
The customer reported the device was discarded.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.G9: exemption number e2019001-permits numbering sequence to begin with 10000, to avoid duplication of report numbers due to process transition.There may be gaps in numbering for reports submitted during the transition period.
 
Event or Problem Description
It was reported that this was a percutaneous intervention, treating the proximal-mid superficial femoral artery (sfa).An unspecified 80mm stent was implanted.Following, the armada dilatation catheter advanced to the stent for post-dilatation.Once at the stent, the balloon was sticking out of the stent and the balloon was deemed longer than expected.The implanted stent was 80mm and the balloon length was also 80mm and the operator stated the balloon should not have been observed as longer than the stent.The balloon successfully post-dilated the stent without a device issue.There were no adverse patient effects and there was no clinically significant delay.No additional information was provided regarding this issue.
 
Additional Manufacturer Narrative
Exemption number e2019001-permits numbering sequence to begin with 10000, to avoid duplication of report numbers due to process transition.There may be gaps in numbering for reports submitted during the transition period.The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.Based on the information provided, a conclusive cause could not be determined.Per the armada 18 product specification, the workable balloon length measured from distal balloon shoulder to proximal balloon shoulder with balloon inflated to nominal pressure (np) for the 80 mm balloon has a tolerance of -5mm/+3mm.It may be possible that the balloon in question was at the higher end of the specification and therefore appeared oversized; however, this could not be confirmed.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 18 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key8922296
Report Number2024168-2019-11067
Device Sequence Number2396983
Product Code LIT
UDI-Device Identifier08717648208416
UDI-Public08717648208416
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K151317
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2021
Device Catalogue Number1013470-080
Device Lot Number8041941
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/30/2019
Supplement Date Received by Manufacturer08/23/2019
Initial Report FDA Received Date08/22/2019
Supplement Report FDA Received Date09/10/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age74 YR
Patient Weight61
-
-