| Model Number |
CD3231-40Q |
| Medical Device Problem Code |
Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device is included in the battery performance alert advisory issued by abbott on 28 august 2017.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient presented in clinic for routine follow up after receiving a vibratory alert.Upon interrogation, it was noted that elective replacement indicator (eri) and battery performance alert (bpa) were triggered.Following the battery performance alert (bpa) advisory, a bpa was received by the clinician and the device was explanted and replaced.The patient was stable post procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported field event of a battery performance alert (bpa) was confirmed via review of device image.The alert was due to a transient drop in the battery voltage that is consistent with battery depletion associated with lithium clusters.However, since the monthly battery voltage trend and the overall expected longevity of the device was normal, premature battery depletion could not be confirmed.
|
| |
|