Model Number CD3231-40Q
Medical Device Problem Code Battery Problem (2885)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device is included in the battery performance alert advisory issued by abbott on 28 august 2017.
 
Event or Problem Description
It was reported that the patient presented in clinic for routine follow up after receiving a vibratory alert.Upon interrogation, it was noted that elective replacement indicator (eri) and battery performance alert (bpa) were triggered.Following the battery performance alert (bpa) advisory, a bpa was received by the clinician and the device was explanted and replaced.The patient was stable post procedure.
 
Additional Manufacturer Narrative
The reported field event of a battery performance alert (bpa) was confirmed via review of device image.The alert was due to a transient drop in the battery voltage that is consistent with battery depletion associated with lithium clusters.However, since the monthly battery voltage trend and the overall expected longevity of the device was normal, premature battery depletion could not be confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY CRT-D
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
MDR Report Key8923266
Report Number2938836-2019-12718
Device Sequence Number19364964
Product Code NIK
UDI-Device Identifier05414734504393
UDI-Public05414734504393
Combination Product (Y/N)N
Initial Reporter StateAK
Initial Reporter CountryUS
PMA/510(K) Number
P030054
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 11/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2013
Device Model NumberCD3231-40Q
Device Catalogue NumberCD3231-40Q
Device Lot Number3523069
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/29/2019
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 08/08/2019
Supplement Date Received by Manufacturer11/15/2019
Initial Report FDA Received Date08/22/2019
Supplement Report FDA Received Date11/24/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0116-2017
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age38 YR
-
-