| Model Number |
720054-02 |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/23/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient underwent a surgical procedure to replace this spectra malleable penile prosthesis due to poor rigidity and quality of erections preventing intercourse.The existing device was explanted and an inflatable penile prosthesis (ipp) was implanted in its place.No patient complications were reported.The explanted components are expected for return.
|
| |
|
Additional Manufacturer Narrative
|
|
Product investigation: the complaint component was returned and analyzed, and the reported allegation of poor rigidity was not confirmed via product analysis.The spectra cylinders were visually inspected and functionally tested.Both cylinders performed within specifications.The product record review indicated that the product met all in-process specifications prior to leaving bsc and the reported events do not represent a new or unanticipated event.The investigation conclusion code of no problem detected was chosen because a device complaint or malfunction could not be confirmed through product analysis.Based on the results of this investigation, no escalation to ncep/capa/scar is required.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent a surgical procedure to replace this spectra malleable penile prosthesis due to poor rigidity and quality of erections preventing intercourse.The existing device was explanted and an inflatable penile prosthesis (ipp) was implanted in its place.No patient complications were reported.The explanted components are expected for return.
|
| |
|
Search Alerts/Recalls
|