MEDTRONIC MINIMED PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723RNAB; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-723RNAB
Medical Device Problem Code Device Difficult to Program or Calibrate (1496)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/06/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.Device received with intermittent button response due to corroded keypad traces.No button error alarm during testing.Device passed displacement test and self test.
 
Event or Problem Description
The customer reported via phone call that the insulin pump alarmed button error.Customer's blood glucose level was not reported.Customer was advised to discontinue use of the device and revert to a backup plan.The customer was advised that the device would be replaced and agreed to return the product for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723RNAB
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325 1219
MDR Report Key8924816
Report Number2032227-2019-52019
Device Sequence Number11269828
Product Code OYC
UDI-Device Identifier00643169221567
UDI-Public(01)00643169221567
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-723RNAB
Device Catalogue NumberMMT-723RNAB
Device Lot NumberA000290053
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/13/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/21/2019
Initial Report FDA Received Date08/23/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age13 YR
-
-