| Medical Device Problem Codes |
Image Display Error/Artifact (1304); Moisture Damage (1405)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
| |
|
Event or Problem Description
|
|
On 14-aug-2019, the reporter contacted animas, alleging a display (blank screen w/moisture) issue.There was no indication that the device caused or contributed to an adverse event.This complaint is being reported because the issue may impact the user's ability to read some or all of the information on the screen which may result in over or under delivery.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the device has been returned and evaluated by product analysis on 31-oct-2019 with the following findings: during investigation, the pump was powered on and the display was found to be clear, legible and functioning properly.Part of the original complaint of a display issue was unable to be duplicated.A leak test was performed and a leak was identified at the crack in the pump case.The pump was opened and moisture damage was found on the printed circuit board.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.
|
| |
|