|
Model Number 45007 |
Device Problems
Break (1069); Difficult to Advance (2920); Material Deformation (2976)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 07/30/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device evaluated by mfr: returned product consisted of a jetstream xc-2.1 atherectomy catheter.The device was visually examined for any shaft damage.Visual examination notice that there were multiple buckling/kinks.The location of the damage was 18cm to 70cm from the tip.The damage was in multiple areas along the location.Visual examination noticed that the infusion line had burst proximal the buckling/kinks.The location of the burst infusion line was approximately 70.5cm to 75.5cm from the tip.The damage that was notice is consistent with sheath interference during the procedure or by pushing, pulling and torqueing of the device in a tortuous anatomy or a heavily calcified lesion.The device was set-up and functionally tested per the dfu and the device functioned as designed except for the leaking at the burst infusion line.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.
|
|
Event Description
|
It was reported that the infusion line bubbled and popped.A 2.1mm jetstream atherectomy catheter was selected for an atherectomy procedure in the left distal superficial femoral artery (sfa) at the adductor canal.The catheter was prepped per usual and used in blades down mode for the entire procedure.During the procedure, the catheter was slowly advanced over a 0.014 thruway guidewire through the occlusion.The catheter advanced until the adductor canal, where it became difficult to advance.As the physician attempted to advance the device, the infusion line of the catheter began to bubble at the entrance to the sheath.A large bubble formed and popped.The catheter was immediately removed from the patient in normal fashion.The procedure was completed with follow up angioplasty and two additional stents placed in the sfa.There were no patient complications.
|
|
Manufacturer Narrative
|
Device evaluated by mfr: returned product consisted of a jetstream xc-2.1 atherectomy catheter.The device was visually examined for any shaft damage.Visual examination notice that there were multiple buckling/kinks.The location of the damage was 18cm to 70cm from the tip.The damage was in multiple areas along the location.Visual examination noticed that the infusion line had burst proximal the buckling/kinks.The location of the burst infusion line was approximately 70.5cm to 75.5cm from the tip.The damage that was notice is consistent with sheath interference during the procedure or by pushing, pulling and torqueing of the device in a tortuous anatomy or a heavily calcified lesion.The device was set-up and functionally tested per the dfu and the device functioned as designed except for the leaking at the burst infusion line.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.
|
|
Event Description
|
It was reported that the infusion line bubbled and popped.A 2.1mm jetstream atherectomy catheter was selected for an atherectomy procedure in the left distal superficial femoral artery (sfa) at the adductor canal.The catheter was prepped per usual and used in blades down mode for the entire procedure.During the procedure, the catheter was slowly advanced over a 0.014 thruway guidewire through the occlusion.The catheter advanced until the adductor canal, where it became difficult to advance.As the physician attempted to advance the device, the infusion line of the catheter began to bubble at the entrance to the sheath.A large bubble formed and popped.The catheter was immediately removed from the patient in normal fashion.The procedure was completed with follow up angioplasty and two additional stents placed in the sfa.There were no patient complications.
|
|
Search Alerts/Recalls
|
|
|