• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. PROLENE SUTURE 2 X 18"(45CM) 7-0 BLU; SUTURE, NONABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. PROLENE SUTURE 2 X 18"(45CM) 7-0 BLU; SUTURE, NONABSORBABLE, SYNTHETIC Back to Search Results
Catalog Number EPM8736
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/01/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint #: (b)(4).The review of the batch manufacturing records was conducted, and the batch met all finished goods release criteria.Summary: three empty opened straight pack and an unopened sample received for evaluation.The product code is multi-strand and contains two strands per packet.During the visual inspection of the sample, no defects were found on the package.The sample was opened, and the swage and attachment area were noted to be as expected.The suture was dispensed without problems and examined along of the strand and no defects, damaged or suture breakage were found.A functional test was performed and the tensile strength force was above the minimum requirement.Per the condition of the representative sample, no performance - breakage suture was found and the tested sample met the finished goods requirements.Representative samples returned to support investigation of unknown quantity of device issues captured in reports 2210968-2019-86071 and 2210968-2019-86072.Analysis results have been included in initial reports for each.
 
Event or Problem Description
It was reported that a patient underwent an unknown procedure in 2019 and the suture was used.It was reported that the suture kept breaking during the anastomosis.It was not reported how the procedure was completed.There were no patient consequences reported.
 
Additional Manufacturer Narrative
It was reported that a patient underwent a coronary artery bypass grafting/cabg times 2 on (b)(6) 2019 and the suture was used.While physician was suturing an anastomosis of the heart, the suture broke half way around.There were no patient consequences reported.This medwatch report is in response to receipt of voluntary medwatch event report (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROLENE SUTURE 2 X 18"(45CM) 7-0 BLU
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876 0151
MDR Report Key8926383
Report Number2210968-2019-86072
Device Sequence Number242727
Product Code GAW
UDI-Device Identifier10705031203679
UDI-Public10705031203679
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K133356
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2023
Device Catalogue NumberEPM8736
Device Lot NumberMEH216
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/19/2019
Initial Date Received by Manufacturer 07/29/2019
Supplement Date Received by Manufacturer07/29/2019
Initial Report FDA Received Date08/23/2019
Supplement Report FDA Received Date08/23/2019
Patient Sequence Number1
Patient Age65 YR
Patient Weight106
-
-