• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS 1 MULTIFOCAL; MULTIFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS 1 MULTIFOCAL; MULTIFOCAL IOLS Back to Search Results
Model Number ZKB00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 05/21/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the power refraction was off, therefore, the intraocular lens (iol) was explanted.The lens was replaced with the same model, but higher diopter.There was no incision enlargement or any other patient injury.The patient is doing fine post-operatively.No additional information was provided to johnson & johnson surgical vision.
 
Additional Manufacturer Narrative
Device available for evaluation? yes.Returned to manufacturer on: 9/11/19.Device returned to manufacturer ¿ yes.Device evaluation: visual inspection with the unaided eye shows the product is cut in pieces, most probably to make explant possible.Considering the condition of the lens additional analysis is not possible.The complaint cannot be confirmed.Manufacturing record review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured according to specifications.The search revealed that no similar complaint for this production order number has been received.Conclusion: as a result of the investigation there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS 1 MULTIFOCAL
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key8926929
Report Number9614546-2019-00772
Device Sequence Number2048240
Product Code MFK
UDI-Device Identifier05050474560055
UDI-Public(01)05050474560055(17)220415
Combination Product (Y/N)Y
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/10/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/15/2022
Device Model NumberZKB00
Device Catalogue NumberZKB00U0185
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/11/2019
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 07/30/2019
Supplement Date Received by Manufacturer09/17/2019
10/25/2020
Initial Report FDA Received Date08/23/2019
Supplement Report FDA Received Date10/11/2019
11/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR
Date Report Sent to FDA01/01/2005
-
-