| Catalog Number |
657012S |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/31/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device will not be returned.If additional information becomes available, it will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
It was reported that the locking screw was not locked during variax clavicle surgery.Therefore, the re-drilling was done but the screw was not locked.After that, the screw was changed and the operation was completed successfully with no surgical delay or adverse consequences reported.
|
| |
|
Event or Problem Description
|
|
It was reported that the locking screw was not locked during variax clavicle surgery.Therefore, the re-drilling was done but the screw was not locked.After that, the screw was changed and the operation was completed successfully with no surgical delay or adverse consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could not be confirmed, since the device was not returned for evaluation and no other evidence was provided.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
|
| |
|